Features & Benefits
Polyurethane Material
Allows for long-term indwelling time.
Large Drainage Holes
Help maximize drainage.
Unique Design
Catheter is designed to allow unimpeded drainage.
Depth Markers
Depth markers and radiopaque marker band helps with fast and accurate positioning.
Atraumatic Pigtail
Designed to reduce the potential for damaging the duodenum.
Pigtail Straightener
Allows for easy insertion of the stiffening cannulas.
Hydrophilic Coating
Facilitates smooth insertion.
Two Drainage Hole Patterns
Accommodate multiple lesion locations and varying patient anatomy.
MRI INFORMATION
MRI INFORMATION
MR CONDITIONAL
The ReSolve+ Biliary Locking Drainage Catheter is MR Conditional
Non-clinical testing has demonstrated that the ReSolve+ Biliary Locking Drainage Catheter is MR Conditional. This device can be scanned safely in a patient, immediately after placement under the following conditions:
- Static Magnetic Field of 3 Tesla or less
- Maximum spatial gradient magnetic field of 4,000 gauss/cm (40 T/m)
- Maximum MR system reported, whole body averaged specific absorption rate (SAR) of 4-W/kg in the First Level Controlled Operating Mode of operation for the MR system
MRI-RELATED HEATING
Under the scan conditions defined above, the ReSolve+ Biliary Locking Drainage Catheter is expected to produce a maximum temperature rise of 2.3°C after 15 minutes of continuous scanning.
ARTIFACT INFORMATION
The maximum artifact size as seen on the gradient echo pulse sequence at 3-Tesla extends approximately 2-mm relative to the size of the shape of the ReSolve+ Biliary Locking Drainage Catheter.