Features & Benefits
MRI INFORMATION
MR CONDITIONAL
The ReSolve+ Biliary Locking Drainage Catheter is MR Conditional
Non-clinical testing has demonstrated that the ReSolve+ Biliary Locking Drainage Catheter is MR Conditional. This device can be scanned safely in a patient, immediately after placement under the following conditions:
- Static Magnetic Field of 3 Tesla or less
- Maximum spatial gradient magnetic field of 4,000 gauss/cm (40 T/m)
- Maximum MR system reported, whole body averaged specific absorption rate (SAR) of 4-W/kg in the First Level Controlled Operating Mode of operation for the MR system
MRI-RELATED HEATING
Under the scan conditions defined above, the ReSolve+ Biliary Locking Drainage Catheter is expected to produce a maximum temperature rise of 2.3°C after 15 minutes of continuous scanning.
ARTIFACT INFORMATION
The maximum artifact size as seen on the gradient echo pulse sequence at 3-Tesla extends approximately 2-mm relative to the size of the shape of the ReSolve+ Biliary Locking Drainage Catheter
Powerful Drainage
Our ReSolve Biliary Locking Drainage Catheter is designed for FAST, UNIMPEDED DRAINAGE and offers up to 4X the drainage area, with multiple LARGE HOLES and different size configurations. Additional benefits include depth markers and a radiopaque marker band for FAST AND ACCURATE POSITIONING, an atraumatic pigtail designed to reduce the potential for damaging the duodenum, and its polyurethane material allows for long-term indwelling time.
- Two drainage hole patterns to accommodate multiple lesion locations and varying patient anatomy
- Unique locking mechanism to prevent movement.
- Repositioning tool to unlock and reposition catheter.
- Radiopaque marker band helps with fast and accurate positioning.
The different hole patterns of the ReSolve Biliary Locking Drainage Catheter provide targeted drainage of both distal lesions (RBDC option) and proximal lesions (RBC option).